AccuSense
TM
Model ASM
High Performance Pressure Transducer
Installation Guide
DESCRIPTION
The AccuSense
TM
Model ASM pressure transducer is a high performance pres-
sure transducer designed for accurate, reliable pressure measurements. It has
a high level analog output signal, excellent stability, and secure calibration
which makes it ideal for high performance industrial, laboratory, and engine
test cell applications.
1.0 GENERAL INFORMATION
Every Model ASM has been tested and calibrated before shipment. Specic
performance specications are shown on page 3 of this Guide. Setra Systems
ASM pressure transducers sense gauge, absolute, or vacuum gauge pres-
sure and convert this pressure dierence to a proportional high level analog
output. Voltage outputs of 0 to 5VDC or 0 to 10VDC, and current output of 4
to 20 mA are oered.
2.0 MECHANICAL INSTALLATION
2.1 Media Compatibility
Model ASM transducers are designed for use with gases and liquids compat-
ible with 17-4PH stainless steel.
2.2 Environment
The operating temperature limits of the ASM are -40°C to +85°C (-40°F to
+185°F)
The compensated temperature range is -20°C to +60°C (-4°F to +140°F)
2.3 Pressure Fittings
Available pressure ttings are given in table below:
Pressure Fitting Code Fitting Description
1F 1/8”-27 NPT Female
1M 1/8”-27 NPT Male
2F 1/4”-18 NPT Female
2M 1/4”-18 NPT Male
J7 7/16”- 20 SAE Male
Vista de página 0
1 2 3 4 5 6 7 8

Resumo do Conteúdo

Página 1 - Installation Guide

1AccuSenseTM Model ASM High Performance Pressure TransducerInstallation Guide DESCRIPTIONThe AccuSenseTM Model ASM pressure transducer is a high pe

Página 2 - Viewed from Front

22.4 Installation of Pressure FittingsYour transducer is designed for the most accurate operation when subjected to pressures within the designated pr

Página 3 - Electrical Data

3The current ows into the +EXC terminal and returns back to the power sup-ply through the -EXC terminal (See Diagram 1). Note: The +SIG Out and -SIG

Página 4

44.2 Zero Adjustment While applying zero pressure, zero oset may be adjusted by pressing the send button to tare zero. If ne adjustment is needed on

Página 5 - Performance Data

5Pressure RangesStandard Code “00”High Over pressure-Option Code “01”Full Scale Range(PSI)BurstPressure*(PSI)Proof Pressure**(PSI)High Proof Pressure

Página 6 - Accuracy Code

6Zero Oset Position Eect <0.05%/g (Ranges ≥ 100 psi) <0.1%/G (Ranges ≤ 50 psi) Unit factory calibrated in vertical position (Pressure Port dow

Página 7 - ORDERING INFORMATION

7Model ASM ORDERING INFORMATIONCode all blocks in table.Note: Please reference AccuSenseTM Brochure or Setra.com website for latest available congura

Página 8 - SS-ASM-Rev C 07/2014

86.0 RETURNING PRODUCTS FOR REPAIRPlease contact a Setra application engineer (800-257-3872, 978-263-1400) before returning unit for repair to review