Installation Guide1-800-257-3872 Toll Free1-978-264-0292 Faxwww.setra.com Web SiteModel 223Ultra-High Purity Flow-ThroughPressure TransducerISO✓Certif
2Setra Model 223Ultra High Purity Flow-Through Pressure Transducer1.0 GENERAL INFORMATIONEvery Model 223 has been tested and calibrated before shipmen
When installing units with Face Seal Fittings:a) Align piping system to transducer connectionsb) Hand tighten nutsc) Torque nuts by placing wrenches o
44.0 CALIBRATIONThe 223 transducer is factory calibrated to the specific input pressurerange vs. output voltage or current and should require no fie
54.1 Voltage Output Zero AdjustmentWhile monitoring the voltage between the positive output (+OUT) andcommon (COM), and with the pressure port open t
BAR COMPOUND CURRENTRANGE OUTPUT (mA)-1 to 1.7 9.92-1 to 3.4 7.63-1 to 7 6.00-1 to 17 4.88-1 to 35 4.44-1 to 70 4.22-1 to 200 4.074.3 Current Output
76.0 RETURNING PRODUCTS FOR REPAIRPlease contact a Setra application engineer (800-257-3872, 978-263-1400) beforereturning unit for repair to review i